Correlation of Early Postpartum Anal/Faecal Incontinence with Endoanal Ultrasound, Digital Rectal Examination, Machine Learning-supported Electric Impedance Spectroscopy, and High-Resolution Anorectal Manometry Results: Exploratory Results from a Prospective, Comparative, Multicentre Clinical Study

FERNANDO R1, HERMAN H2, UCHMAN-MUSIELAK M3, DOSEDLA E4, GONZALEZ DIAZ E5, JANKU P6, ATHANASIOU S7, BAEKELANDT J8, DZIKI A9, FRANCO R10, GRZESIAK M11, RATTO C12, OSZUKOWSKI P11, RIZK D13, SALVATORE S14, STUART A15, DE TAYRAC R16, SPINELLI A17, MLYNCZAK M18, IWANOWSKI P18, KORZENIEWSKI K18, ROSOL M18, BORYCKA K18

Research Type

Clinical

Abstract Category

Anorectal / Bowel Dysfunction

Gold Pass video coming soon
Find out more

Abstract 307
Pregnancy and Pelvic Floor Disorders
Scientific Podium Short Oral Session 29
Friday 25th October 2024
16:15 - 16:22
Hall N105
Anal Incontinence New Devices Female
1. Imperial College Healthcare NHS Trust, London, 2. Institute for the Care of Mother and Child, Prague, Czech Republic, 3. Solec Hospital, Warsaw, Poland, 4. Univerzita Pavla Jozefa Šafárika, Košice, Slovakia, 5. Complejo Asistencial Universitario de Leon, Spain, 6. Masaryk University, Brno, Czech Republic, 7. Kapodistrian University of Athens, Greece, 8. Imelda Hospital, Bonheiden, Belgium, 9. Medical University of Lódz, Poland, 10. Fondazione Policlinico Universitario A. Gemelli IRCCS, Rome, Italy, 11. Institute of Polish Mother's Health Centre, Lódz, Poland, 12. Fatebenefratelli Gemelli Isola Hospital Rome, Italy, 13. Arabian Gulf University, Bahrain, 14. University Vita e Salute, IRCCS San Raffaele, Milan, Italy, 15. Department of Obstetrics and Gynaecology, Lund, and Department of Gynaecology, Helsingborg, Sweden, 16. CHRU Carémeau, Nîmes, France, 17. Humanitas University, Milan, Italy, 18. OASIS Diagnostics SA, Warsaw, Poland
Presenter
Links

Abstract

Hypothesis / aims of study
Anal incontinence (AI), a debilitating outcome of obstetric anal sphincter injuries (OASIs), significantly impacts quality of life of women affected by the latter. The early diagnosis of OASIs is particularly challenging as anorectal symptoms, in particular AI, can be subtle or not yet developed shortly after delivery, as indicated by largely retrospective literature. This discrepancy underscores the urgent need for robust, prospective data to leverage both diagnosis and management of OASIs. Addressing this critical gap, this study evaluated the correlation between early postpartum AI symptoms and results of diagnostic tests addressing the integrity of anal sphincter complex or its function, the former including the novel Machine Learning-supported electric impedance spectroscopy (ONIRYTM OASIS Diagnostic) test).
Study design, materials and methods
This exploratory analysis was performed within a prospective, comparative clinical study on detection of OASI conducted at four European labour wards and one outpatient maternal care clinic in 152 women shortly (up to 5 weeks) after vaginal delivery, enriched for women with OASI (n=60) as determined by endoanal ultrasound (EAUS) (the primary results thereof pending publication). AI symptoms were analysed using Wexner score, not earlier than 3 days postpartum. The correlation analysis for Wexner total score was evaluable and performed using Spearman's correlation tests for 54, 54, 53, and 46 women with OASI and 62, 62, 62, and 55 women without OASI for results of digital rectal examination (subjectively suggestive of OASI), electric impedance spectroscopy (ONIRYTM test result compatible with OASI), EAUS (grades 3 and 4 per OASIS classification), and high-resolution anorectal manometry (major finding per London Classification by International Anorectal Physiology Working Group).
Results
The results of the correlation analysis are shown in the table 1. 
		
The main clinical study had two parts: the clinical conduct and the in silico part; during the latter the Machine Learning model was trained using 10 iterations of 10-fold cross-validation. Therefore, in case of ONIRYTM test, the correlation coefficient was calculated as a mean value obtained from 10 iterations of each cross-validation, thus there is no single corresponding p-value; the one presented in the table is the highest of the ten (still being statistically significant).
Interpretation of results
The analysis revealed significant correlations between the Wexner total score and all four diagnostic methods evaluated. Notably, the strongest correlation was with the physical digital rectal examination whereas the correlation coefficient for the other 3 methods was less strong, including the anorectal manometry. This may be due to the fact that the digital rectal examination, known for its limited sensitivity for detection of OASI, is able to reliably detect mostly the cases with markedly reduced anal sphincter tonus, which are compatible with FI. The somewhat lower correlation with anorectal manometry abnormal results, despite the latter being the gold standard method for anal sphincter dysfunction, may illustrate the latency between the actual, early dysfunction measurable by this highly sensitive and specific method and the development of clinically meaningful FI symptoms. The correlation with EAUS and ONIRYTM test, both statically evaluating the integrity of perianal structures, at a comparably lower level, may also illustrate the latency between the injury (OASI) and development of dysfunction (FI); the latter frequently not being immediate.
Concluding message
This exploratory study provides data indicating that the early postpartum AI symptoms are suggestive of OASI but with only moderate level of correlation. Knowing the low sensitivity of digital rectal examination for detection of OASI as such, this data underscores the necessity for a multifaceted diagnostic strategy considering both traditional and innovative methods to address the complexities of OASI detection effectively. While EAUS is of marginal accessibility at obstetric settings as a screening tool for OASI, explored for this purpose should be the novel, more accessible methods such Machine Learning-supported electric impedance spectroscopy provided with the ONIRYTM test.
Figure 1 Table 1
Disclosures
Funding The study was financed by the European Union as part of the Fast Track program, conducted in Poland by the Polish National Centre for Research and Development (POIR.01.01.01-00-0726/18) Clinical Trial Yes Public Registry No RCT No Subjects Human Ethics Committee Eticka komise, UPMD, Podolské nábreží 157, 147 10 Praha, Czech Rep. Eticka komise, Fakultni Nemocnice Brno, Jihlavska 20, 625 00 Brno, Czech Rep. CEIm de las Áreas de Salud de León y del Bierzo Complejo Asistencial Universitario de León Calle Altos de nava, 24001 León, Spain Komisja Bioetyczna przy Okregowej Izbie Lekarskiej w Warszawie, ul. Pulawska 18, 02-512 Warszawa, Poland Eticka komisia pri Nemocnica AGEL Košice-Šaca a.s., Lúcna 57, 040 15 Košice-Šaca, Slovakia Helsinki Yes Informed Consent Yes
Citation

Continence 12S (2024) 101649
DOI: 10.1016/j.cont.2024.101649

24/11/2024 01:59:05